Journal Information
Vol. 56. Issue 8.
Pages 540 (August 2020)
Share
Share
Download PDF
More article options
Vol. 56. Issue 8.
Pages 540 (August 2020)
Letter to the Editor
Full text access
Impact of Using a Novel Gastric Feeding Tube Adapter on Patient's Comfort and Air Leaks During Non-invasive Mechanical Ventilation
Impacto del uso de un nuevo adaptador de sonda de alimentación gástrica en la comodidad del paciente y en las fugas de aire durante la ventilación mecánica no invasiva
Visits
1653
Alejandro Úbedaa,
Corresponding author
alubeda@gmail.com

Corresponding author.
, Irene Fernándeza, Antonio M. Esquinasb
a Critical Care Unit, Hospital Punta de Europa, Carretera de Getares, s/n, Algeciras, Cádiz, Spain
b Critical Care Unit, Hospital Morales Meseguer, Av. Marqués de los Velez s/n, Murcia, Spain
This item has received
Article information
Full Text
Bibliography
Download PDF
Statistics
Full Text
Dear Editor,

We share the concerns expressed by Quintero et al.1 on comfort and air leakage during non-invasive mechanical ventilation (NIMV) with the need of using oral or naso-oral probes. The authors designed their device aiming to decrease air leaks to improve patient comfort, which is related to success of NIMV.

After reading the article, we found several discussion points.

First. We think that the selection of patients included as their own controls is a success.

Second. Quintero et al.1 included 196 patients but there is no flowchart explaining how many patients met the inclusion criteria and how many were excluded and the reasons.

Third. To asses comfort, they used a qualitative ordinal scale with only 5 options to choose from. When using that type of scales, there could be a tendency of regression toward the central value, so that measurement bias could appear.

Fourth. It is not explained if patients begin with NIVM using the adapter or the conventional tube. If a patient first receives the most uncomfortable one, when it is changed to the comfortable one, his perception of comfortability may be higher. We believe that this matter should have been addressed.

Fifth. The authors estimated a sample size of 191 patients and they included 196 patients in the study but comfort was evaluated only in 99 patients with Glasgow Coma Scale of 15, so sample size was not reached for one of the outcome of interest.

Sixth. Respiratory and hemodynamic parameters during NIMV, such as inspiratory and expiratory tidal volumes, oxygen saturation and respiratory rate, were better in the NIMV with tube adaptor group, but it is difficult to know if it was the result of TA-NIMV alone or the sum of both therapies.

Seventh. Therapy was performed according to individual needs, so patients could have received different therapies and that heterogeneity could make patients non-comparable.

Quintero et al.1 concluded that the use of their tube adaptor to adjust NIMV mask interfaces in patients in whom oro or naso-enteral tubes were in place, significantly reduced air leaks and improved subjective comfort perceived by the patient during NIMV support.

As there has been no important changes in recommendations published within the past 15 years regarding the use of NIMV for various forms of respiratory failure 2, we believe that proposals like Quintero's gastric feeding tube adaptor may improve the assistance of patients with respiratory failure during NIMV, even though further studies should be performed to correct biases.

References
[1]
O.I. Quintero, A.I. Sánchez, P.A. Chavarro, I.C. Casas, G.A. Ospina.
Impact of using a novel gastric feeding tube adaptor on patient's comfort and air leaks during non-invasive mechanical ventilation.
Arch Bronconeumol, 56 (2020), pp. 353-359
[2]
B. Rochwerg, L. Brochard, M.W. Elliott, D. Hess, N.S. Hill, S. Nava, et al.
Official ERS/ATS clinical practice guidelines: noninvasive ventilation for acute respiratory failure.
Copyright © 2020. SEPAR
Archivos de Bronconeumología
Article options
Tools

Are you a health professional able to prescribe or dispense drugs?