Review
Pulse oximetry: Principles and limitations

https://doi.org/10.1016/S0735-6757(99)90019-0Get rights and content

Abstract

The pulse oximeter has become an essential tool in the modern practice of emergency medicine. However, despite the reliance placed on the information this monitor offers, the underlying principles and associated limitations of pulse oximetry are poorly understood by medical practitioners. This article reviews the principles of pulse oximetry, with an eye toward recognizing the limitations of this tool. Among these are performance limitations in the settings of carboxyhemoglobinemia, methemoglobinemia, motion artifact, hypotension, vasoconstriction, and anemia. The accuracy of pulse oximetry is discussed in light of these factors, with further discussion of applications for pulse oximetry in emergency medicine, including both oximetric and plethysmographic operation. The pulse oximeter is an invaluable instrument for emergency medicine practice, but as with any test the data it offers must be critically appraised for proper interpretation and utilization.

References (76)

  • GD Jay et al.

    Pulse oximetry is accurate in acute anemia from hemorrhage

    Ann Emerg Med

    (1994)
  • C Cahan et al.

    Agreement between noninvasive oximetric values for oxygen saturation

    Chest

    (1990)
  • AL Ries et al.

    Skin color and ear oximetry

    Chest

    (1989)
  • J Jones et al.

    Continuous emergency department monitoring of arterial saturation in adult patients with respiratory distress

    Ann Emerg Med

    (1988)
  • JR Mateer et al.

    Continuous pulse oximetry during emergency endotracheal intubation

    Ann Emerg Med

    (1993)
  • DG Levinsohn et al.

    The Allen's test: Analysis of four methods

    J Hand Surg [Am]

    (1991)
  • B Graham et al.

    Pulse oximetry for vascular monitoring in upper extremity replantation surgery

    J Hand Surg [Am]

    (1986)
  • LR Rodriguez et al.

    A study of pediatric house staff's knowledge of pulse oximetry

    Pediatrics

    (1994)
  • JF Kelleher

    Pulse oximetry

    J Clin Monit

    (1989)
  • RF Grace

    Pulse oximetry

    Gold standard or false sense of security?

    Med J Aust

    (1994)
  • KK Tremper et al.

    Pulse oximetry

    Anesthesiology

    (1989)
  • MW Wukitsch et al.

    Pulse oximetry: analysis of theory, technology, and practice

    J Clin Monit

    (1988)
  • JW Severinghaus et al.

    Accuracy of response of six pulse oximeters to profound hypoxemia

    Anesthesiology

    (1987)
  • D Thrush et al.

    Accuracy of pulse oximetry during hypoxemia

    South Med J

    (1994)
  • SJ Barker et al.

    The effect of carbon monoxide inhalation on pulse oximetry and transcultaneous PO2

    Anesthesiology

    (1987)
  • ST Anderson et al.

    Benzocaine-induced methemoglobinemia in an adult: accuracy of pulse oximetry with methemoglobinemia

    Anesth Analg

    (1988)
  • JB Eisenkraft

    Pulse oximeter desaturation due to methemoglobinemia

    Anesthesiology

    (1988)
  • SJ Barker et al.

    Effects of methemoglobinemia on pulse oximetry and mixed venous oximetry

    Anesthesiology

    (1989)
  • JV Anderson

    The accuracy of pulse oximetry in neonates: effects of fetal hemoglobin and bilirubin

    J Perinatol

    (1987)
  • R Ramanathan et al.

    Pulse oximetry in very low birth weight infants with acute and chronic disease

    Pediatrics

    (1987)
  • AP Harris et al.

    Absorption characteristics of human fetal hemoglobin at wavelengths used in pulse oximetry

    J Clin Monit

    (1988)
  • J Pologe et al.

    Effects of fetal hemoglobin on pulse oximetry

    J Perinatol

    (1987)
  • SN Beall et al.

    Jaundice, oximetry, and spurious hemoglobin desaturation

    Anesth Analg

    (1989)
  • F Veyckemans et al.

    Hyperbilirubinemia does not interfere with hemoglobin saturation measured by pulse oximetry

    Anesthesiology

    (1989)
  • MR Kessler et al.

    Spurious pulse oximeter desaturation with methylene blue injection

    Anesthesiology

    (1986)
  • MS Scheller et al.

    Effects of intravenously administered dyes on pulse oximetry readings

    Anesthesiology

    (1986)
  • A Sidi et al.

    Methylene blue and indocyanine green artifactually lower pulse oximetry readings of oxygen saturation

    J Clin Monit

    (1987)
  • L Hanowell et al.

    Ambient light affects pulse oximeters

    Anesthesiology

    (1987)
  • Cited by (369)

    • A comparative study of common steps in video-based remote heart rate detection methods

      2022, Expert Systems with Applications
      Citation Excerpt :

      Therefore, it has been of interest for many decades to find better methods for heart rate detection. The invention of pulse oximeters was an important step towards minimising such human errors, with pulse oximeters being widely used today (Sinex, 1999). One limitation of pulse oximeters is that they must be attached to the human body.

    View all citing articles on Scopus
    View full text