Spirometry is the main diagnostic procedure for most respiratory diseases. This document sets out the principal recommendations for performing and interpreting the test in adult and pediatric populations, in accordance with current evidence-based guidelines and standards. It is important to be familiar with the indications and contraindications of the test, as well as the technical requirements of the equipment and the facilities where it is performed. The personnel responsible for conducting the test must have appropriate training in order to obtain maneuvers of sufficient quality for subsequent interpretation. The quality grade of each test must be determined before interpretation. Spirometric tests should be interpreted using updated reference values appropriate for each patient group, and in addition to assessing whether the measured values are within normal limits, spirometry can provide information suggestive of specific patterns indicative of airway disease, parenchymal lung disease, or even disorders of the chest wall. Performing spirometry after administration of a bronchodilator is also often required for the diagnosis of certain respiratory diseases, and standardization of this procedure is essential in order to interpret the changes in spirometric values after bronchodilation.
Spirometry is the main diagnostic test for assessing lung function in most respiratory diseases. It is considered a basic test that should be performed in pulmonary function laboratories in specialized care, in primary care, and sometimes in other healthcare settings [1]. Since the 2013 guidelines of the Spanish Society of Pulmonology and Thoracic Surgery (SEPAR) [1], changes have been introduced in the recommendations for standardizing the performance and interpretation of the test, mainly driven by the Global Lung Function Initiative (GLI) [2], a clinical research collaboration that brings together physiologists, pulmonologists, epidemiologists, and statisticians, with the support of the main international respiratory disease societies. This document compiles the recommendations for performing this test based on a review of the available scientific literature. Recommendations without sufficient scientific evidence have been agreed upon by consensus among the experts who participated in the literature review and drafting of this document.
Applications of spirometry. Indications and contraindications (Table 1)Indications or applications of spirometry [1,3–6]Spirometry is used to help establish a clinical diagnosis, quantify the severity of pulmonary function impairment, assess the response to treatments, and monitor disease course and patient prognosis.
Indications and contraindications of spirometry.
| Indications | |
|---|---|
| Category | Indications |
| Diagnosis | Evaluate symptoms, signs, or abnormal laboratory test results |
| Measure the physiological effect of disease or disorder | |
| Evaluate individuals at risk of lung disease | |
| Assess preoperative risk | |
| Assess prognosis | |
| Monitoring | Evaluate response to therapeutic intervention |
| Monitor disease progression | |
| Monitor exacerbations and recovery | |
| Monitor adverse effects of exposure to harmful agents | |
| Observe adverse reactions to drugs with known pulmonary toxicity | |
| Evaluation of disability/deterioration | Evaluate patients in rehabilitation programs |
| Assess risks as part of insurance evaluation | |
| Evaluate for legal reasons | |
| Other | Research and clinical trials |
| Epidemiological surveys | |
| Derivation of reference equations | |
| Pre-employment and monitoring of lung health for at-risk occupations | |
| Evaluation of health before risky physical activities | |
| Relative contraindications | |
|---|---|
| Category | Contraindications |
| Cardiovascular | Acute myocardial infarction within the last week |
| Systemic hypotension or severe hypertension | |
| Significant arrhythmia | |
| Uncompensated heart failure | |
| Uncontrolled pulmonary hypertension | |
| Acute cor pulmonale | |
| Unstable pulmonary embolism | |
| History of syncope related to forced maneuvers/cough | |
| Neurological | Cerebral aneurysm |
| Recent brain surgery (last 4 weeks) | |
| Recent concussion with symptoms | |
| Recent eye surgery (last week) | |
| ENTa/sinuses/middle ear | Recent sinus or middle ear surgery or infection (last week) |
| Active or suspected respiratory infection, including tuberculosis | |
| Thoracic/abdominal | Presence of pneumothorax |
| Recent thoracic surgery (last 4 weeks) | |
| Recent abdominal surgery (last 4 weeks) | |
| Late pregnancy | |
| Infectious | Conditions predisposing to transmission of infections (hemoptysis, secretions, oral lesions, oral bleeding) |
| Other | Confusion, dementia, or inability to cooperate |
It is considered a low-risk and well-tolerated procedure. However, it may cause discomfort and potential harm. Harm is usually related to increased intrathoracic, intra-abdominal, and intracranial pressure during the forced maneuver. This maneuver requires maximal, forced inspiratory and expiratory effort, hence the possibility of complications in the situations described below. It may affect abdominal and thoracic organs, increase myocardial oxygen demand, cause changes in venous return and systemic blood pressure, and modify chest wall and lung expansion. Traditionally, thoracic and abdominal aneurysm were also considered relative contraindications for spirometry. Limited data did not show adverse events in aortic aneurysms measuring 5–13cm and in thoracic aortic aneurysms measuring 5–8cm [7]. These contraindications are considered relative, and the decision to perform spirometry is made by the prescribing physician based on an individual risk–benefit assessment for the specific patient. Potential contraindications should be included in the spirometry request form.
The setting in which spirometry is performed should influence the decision: performing spirometry in the presence of a relative contraindication is not the same in a facility with access to emergency care as in one without it. In the latter case, referral to a specialized center may be necessary.
Since spirometry requires cooperation between the patient and the operator, it is sometimes not possible to perform it in confused, demented, or severely ill patients. Spirometry should be discontinued if the patient experiences pain during the maneuver, has a syncopal episode, or if FEV1 falls by ≥20% from baseline [8].
Nevertheless, complications related to spirometry are very rare, and it is considered a very safe test when the aforementioned situations are taken into account. The incidence of complications is 0.6 per 1000 spirometry tests. The most frequent complications are cardiopulmonary events, particularly syncope.
Recommended equipmentThe physical space where pulmonary function tests are performed should be quiet, comfortable, and adequately ventilated (naturally or with air renewal between 8 and 15 times per hour). It should be acoustically isolated, with non-porous and easy-to-clean materials. Sufficient space (minimum 2.5m×3m) is required to accommodate the patient and technician, including stretchers for supine examinations. Essential devices include spirometers, computers, sinks, stadiometers, scales, thermometers, hygrometers, barometers, measuring tapes, disposable mouthpieces and nose clips, bronchodilators, and appropriate filters. A specific area for cleaning and disinfecting materials is required [1,3,8]. Equipment maintenance is crucial for proper functioning and is divided into preventive maintenance (planned, performed by laboratory staff) and corrective maintenance (after failure, performed by specialized technicians or trained personnel). It is mandatory to follow the manufacturer's instructions and keep detailed records documenting maintenance actions, calibration, verification, detected problems, and quality controls. These records must be kept according to institutional regulations. All new equipment must be verified and validated before clinical use. Quality control ensures reliable and accurate results. It includes daily calibration, verification of volume and flow accuracy, periodic use of biological controls, and supervision of technician performance. All spirometers must comply with the standards of ISO 26782:2020 [9]. The following are the minimum specifications that a spirometer must meet (Tables 2 and 3) [4,10,11].
Additional technical requirements of the spirometer.
| Requirement | Specification |
|---|---|
| Volume accuracy | Maximum allowed error of 2.5% of measured value or 0.05L, whichever is greater |
| Peak flow accuracy | Maximum allowed error of 10% of measured value or 0.33L/s (20L/min), whichever is greater |
| Flow resistance | Maximum dynamic resistance of 0.5cm H2O/L/s at a flow of 12L/s |
| Maximum measurable volume | Capacity to measure volumes up to 10L |
| Recording time | Capacity to record a minimum expiratory time of 15s |
| Linearity | Maximum allowed variation of 2.5% across the measurement range |
| Repeatability | Maximum standard deviation of 2.5% between repeated measurements |
| Expiratory impedance | Maximum allowed resistance of 0.5cm H2O/L/s at a flow of 12L/s |
| Calibration | Must be performed using a 3-L calibration syringe with a precision of 0.5% of total volume |
| Sampling frequency | Minimum of 100Hz with a resolution of at least 12 bits |
| Curve visualization | Real-time visualization of flow–volume (F–V) and volume–time (V–T) curves |
| Marking and documentation | Units in liters, increments no greater than 0.1L, numbering at intervals not exceeding 1.0L |
| Electrical and mechanical safety | Compliance with IEC 60601-1:2005 standard |
| Biological compatibility | Materials in contact with the patient must be biocompatible |
| Cleaning and disinfection | Allow effective cleaning, sterilization, and disinfection procedures |
| Operating environmental conditions | Temperature from 0°C to 45°C and compensation for barometric pressure and humidity |
| Updated reference values | Use GLI reference equations for all ages 3–95 years [12] |
Additional technical requirements for spirometer in pediatrics.
| Dead space of the equipment | Additional volume (mouthpiece, filter, adapters) must be less than 100mL to minimize impact on measurement, especially in small children. |
| Incentive software | Incorporation of software with visual incentives or interactive games to facilitate child cooperation during the test. |
Prevention of infections in pulmonary function laboratories is essential, as tests such as spirometry can generate aerosols, especially during procedures like bronchodilator testing, increasing the risk of contagion. Therefore, strict hygiene measures are required [13–15]. All centers must implement protocols including handwashing before and after each test, use of masks adapted to the biological risk of each situation, availability of hydroalcoholic gel, and cleaning routines for surfaces and equipment [5,16,17].
The cleaning protocol should follow Spaulding's classification [18]:
Critical devices: require total sterilization. Semicritical devices (e.g., mouthpieces): require high-level disinfection. Non-critical devices (e.g., nose clips, casings): are cleaned with water and detergent.
The use of disposable materials, such as mouthpieces and nose clips, is recommended. Reusable components must be rigorously cleaned and disinfected. Parts of the spirometer such as tubes and connections should be cleaned weekly and disassembled at the end of the day. The flow head requires special cleaning if filters are not used, including washing with enzymatic detergent and subsequent disinfection or sterilization if indicated. The use of certified disposable antiviral and antibacterial filters is standard practice, as they protect the equipment and reduce the need for disinfection between patients without affecting measurement quality [19–21]. During epidemic or pandemic situations, measures should be intensified: biological risk control, increased cleaning, good ventilation, use of filters, differentiated patient routes, and restriction of companion access [8,22].
Training of personnel performing spirometrySpirometry is a technique that requires specific training for both performing the test and for quality control and equipment maintenance. Pulmonary function laboratories, together with external agents such as respiratory societies, must ensure comprehensive training for personnel who will perform spirometry. This personnel should have at least a university degree, as training should include basic theoretical knowledge in pulmonology, capacity to perform diagnostic procedures, and specific practical training. Independent performance of spirometry requires an initial period to get to know the equipment and observe the technique, followed by a closely supervised initial practice stage and finally a tutored practice period. The period during which the trainee is considered autonomous may vary, but in the 2013 spirometry guidelines of the Spanish Society of Pulmonology and Thoracic Surgery (SEPAR), it was established as at least 3 months [1]. Currently, after the initial observation period, it seems reasonable to require a minimum number of correctly performed and interpreted tests. Based on previous training experiences, this number could be set at least 30 spirometries [23], maintaining ongoing training and the possibility of resolving doubts in the following months. The 2019 American Thoracic Society (ATS)/European Respiratory Society (ERS) guidelines [5] establish that training for personnel performing spirometry must include knowledge of the procedure, acceptability and repeatability criteria, recording and interpretation of results, quality standards and good clinical practice, as well as correct patient preparation, including possible administration of bronchodilator medication in indicated tests. Continuous refresher training should be maintained, both within laboratories and by external agents such as ERS [24] or SEPAR [25], which have among their main objectives promoting and supporting the training of various professionals, providing accessible theoretical-practical training complementary to that learned in laboratories, as well as a way to stay updated on respiratory pathophysiology and lung function.
Procedure in adult patients: forced, slow, and supine spirometryPreparation of the patient and equipmentPrevious instructionsThe patient will receive information about the purpose and characteristics of spirometry. The request form will include written recommendations prior to the test (Tables 4 and 5).
Recommendations for patients prior to spirometry.
| Do not come fasting but avoid heavy meals before the test. |
| Do not ingest stimulants or central nervous system depressants (coffee, alcohol, benzodiazepines) from 8h before if possible. |
| Do not smoke, vape, or use water pipes at least 1 hour before the test. |
| Do not perform vigorous exercise 1 hour before the test. |
| Do not wear tight clothing that limits thoracic/abdominal expansion. |
| Discontinue inhaled medication in indicated cases.a |
Recommended abstinence time according to bronchodilator drug used.
| Bronchodilator medication | Abstinence time |
|---|---|
| SABA | 6h |
| SAMA | 12h |
| LABA | 24h |
| Ultra-LABA | 48h |
| LAMA | 48h |
| Sustained-release theophyllines | 48h |
| Corticosteroids | Not necessary |
| Ensifentrine | 12ha |
SABA: short-acting beta-agonist. SAMA: short-acting anticholinergic/antimuscarinic. LABA: long-acting beta-agonist. LAMA: long-acting anticholinergic/antimuscarinic. Ultra-LABA: long-acting beta-agonist with half-life greater than 24h.
Ensifentrine is a new dual inhaled phosphodiesterase (PDE) inhibitor with combined bronchodilator and anti-inflammatory effects. The elimination half-life of ensifentrine has been reported to be approximately 10–12h [26,27].
Suspension of inhaled medication (if indicated) and absence of acute respiratory symptoms or any other condition listed in the test contraindications must be confirmed. Smoking, occupational, and pharmacological treatment history will be collected. Identifying data (name, medical record, date of birth, biological sex, ethnicity) and anthropometric data (height in centimeters, barefoot, straight back; weight with light clothing, precision 0.5kg) will be recorded. In cases of significant chest or vertebral deformity and/or inability to stand, height may be estimated by arm span [12,28].
Technical preparationA quiet and comfortable environment should be provided, offering the patient 5min of rest before the test. The procedure and instructions should be clearly and concisely explained, adapted to the functional diversity of each person. A mouthpiece with a filter should be placed on the spirometer and its height adjusted. The patient should sit with a straight back, feet supported, and jaw at 90°. If necessary, the test may be performed standing, always recording the position in which it was performed. Lip seal, use of nose clip, and presence of poorly fixed dental prostheses should be checked. In cases of limitations (hemiparesis, neuromuscular diseases, tracheostomy), anatomical rubber mouthpieces, facial masks, or adapters should be considered.
Description of the maneuverForced spirometry: a rapid maximum inspiration to total lung capacity (TLC) will be performed, with a post-inspiratory pause of less than 2s, followed by a forceful and explosive expiration, prolonged until a plateau is reached on the volume–time curve, a flow of less than 25ml/s, or a maximum time of 15s. Then, a new maximum inspiration to TLC should be performed [5]. The maneuver must be continuously supervised by the technician to assess its acceptability. A minimum of 3 acceptable and reproducible maneuvers should be performed according to the criteria described below, and a maximum of 8 maneuvers before selecting the best for interpretation.
Tables 6 and 7 summarize the acceptability and reproducibility criteria for the maneuvers.
Acceptability criteria in adults.
| Back-extrapolated volume (BEV) must be ≤5% of FVC or 100mL. |
| Expiration must be performed continuously and without interruptions. |
| The maximum forced expiration maneuver must stop when the volume change per second is ≤0.025L, for at least 1s (observing a “plateau”). |
| Maximum total forced expiration of 15s with no established minimum time. |
Repeatability criteria in adults.
| Minimum 3 acceptable maneuvers and a maximum of 8. |
| The difference between the two highest values of FVC and FEV1 must be ≤0.150L. |
| The two best maneuvers must not differ from each other by more than 0.150L and the two best FEV1 values must not differ from each other by more than 0.150L or 10% of the highest value (whichever is greater). |
| If FVC is ≤1L, these differences must not be greater than 0.100L or 10% of the highest value (whichever is greater). |
Slow spirometry: slow spirometry, although less used than forced spirometry, can be used in patients where vital capacity (VC) needs to be known and a forced maneuver is not possible (for example, for diffusion testing), or in certain patients with decreased FVC where obstructive origin is suspected [29–31]. The maneuver is performed with the patient breathing at tidal volume, then instructed to perform a deep inspiration to TLC, followed by a deep expiration to reach residual volume.
Supine spirometry: forced spirometry in the supine position is mainly performed in neuromuscular patients, comparing lung function in sitting and supine positions, as its decrease in this position has been related in some studies to diaphragmatic involvement, although its relationship with hypoventilation is not clear [32–35]. The maneuver is performed the same as in the sitting position, but with the patient lying down, and the head may be slightly tilted for greater comfort.
Procedure for performing spirometry in pediatricsPreparationPersonnel working with children must have adequate training, motivation, and experience. It is essential to measure children's height before spirometry. It is not appropriate to use height from previous visits or height reported by parents or caregivers. In children who cannot stand, height should be estimated from ulna length [36]. The child should be placed in the correct posture. As in adult patients, the recommended position is sitting with a straight back. It can also be performed standing, always noting the position adopted [37]. In neuromuscular patients undergoing supine spirometry, this should also be specified. The use of a nose clip is not essential in the forced expiratory maneuver; it is recommended to note whether it was used or not.
TechniqueDepending on the child's abilities and the available equipment or program, one of the two described techniques can be performed:
- 1
Rapid but not forced inspiration, taking all the air until the child reaches total lung capacity; then, insert the mouthpiece held with teeth and close the lips. Without pausing for more than 2s, perform a forceful (rapid and strong) expiration continuously until reaching residual volume. The test should be completed with a strong inspiration to total lung capacity.
- 2
Holding the mouthpiece between the teeth, sealing it with lips, breathe at tidal volume for 2–3 cycles, then rapidly but not forcefully inspire to total lung capacity, and without pausing for more than 2s, perform a forced expiration with maximum effort and speed of all air in the lungs until reaching residual volume. The test should be completed with a strong inspiration to total lung capacity.
The child should be stimulated with words, gestures, and especially body language to encourage maximum inspiration, abrupt initiation of expiration, and sustained expiration until the expiratory plateau is reached. It is highly recommended to use pediatric incentive software that will reinforce specific parts of the maneuver (start, course, or end of forced expiration) [37].
Next, the acceptability and reproducibility criteria for the test in children are specified (Tables 8 and 9).
Acceptability criteria in children under 6 years.
| Rapid start: evidence of initial explosive effort, although the time to peak expiratory flow (PEF) cannot always be measured precisely. |
| No major artifacts: avoid cough in the first 0.5s, re-inhalation, or significant air loss. |
| Minimum expiratory duration 0.5s (in older children objective 1s), although the updated criterion is that it reaches a plateau in expiratory flow that ensures complete expiration (EOFE); new equipment reports whether this is achieved or nota. |
| Flow-volume curve: smooth and without interruptions. |
| Constant effort: without interruptions, with sustained flow until the end. |
Total expired volume (FEV0.5, FEV0.75, FEV1) can be used in place of FEV1 if standard duration is not achieved; in small children, who reach their expiratory plateau earlier due to airway-lung size ratio, adequate expiration may be less than 1s (and it is not that they are not capable of maneuvers with sufficient quality). This can also occur in children under 6 years of age, in whom expiration can be completed before 1s; FEV0.75 values will provide information similar to FEV1. In these cases, interpretation should be careful and contextualized.
Repeatability criteria in children.
| Minimum 2 acceptable maneuvers with no recommended maximum. |
| Repeatability criteria for both FVC and FEV1:• Children older than 6 years: the two best values of FVC and FEV1 must not differ from each other by more than 0.150L• Children younger than 6 years (or FVC less than 1L): differences must not be greater than 0.100L or 10% of the highest value (whichever is greater) |
| The two best must not differ from each other by more than 0.150L and the two best FEV1 values must not differ from each other by more than 0.150L or 10% of the highest value (whichever is greater). |
| If FVC is equal to or less than 1L, differences must not be greater than 0.100L or 10% of the highest value (whichever is greater). |
| In adolescents, adult criteria will be applied. |
Once three acceptable measurements of FVC and FEV1 are achieved, or eight maneuvers are performed (more in children, especially if incomplete maneuvers are performed), the best measurements of these variables will be selected, even if they do not come from the same maneuver. The rest of the parameters and graphs will be obtained from the acceptable maneuver that sums the highest values of FVC and FEV1 (Fig. 1) [36].
Flowchart for applying acceptability and repeatability criteria.
Adapted from Refs. [1,36].
In children ≤6 years, the highest values of FEV0.5 and FEV0.75 will be determined, even if from different maneuvers, collecting the rest of the parameters from the maneuver with the highest sum of FVC+FEV0.5, FVC+FEV0.75, or FVC+FEV1, depending on whether the forced expiratory time (FET) is between 0.5 and 0.74s, 0.75–0.99s, or ≥1s, respectively.
Currently, almost all spirometers incorporate algorithms that automatically evaluate the quality of the maneuver, the best values of the variables, and the curves to be used for test interpretation. However, despite their utility, it is essential that the operator verifies that the selection is appropriate [1].
Classification of measurement qualityA grading system has been proposed to assess the quality of spirometry based on the number of acceptable and reproducible FEV1 and FVC variables (Table 10). Variables of good quality are those of grades A and B, sufficient quality those of grade C, and variables of grades D and lower are not useful for interpretation [1]. The latest ATS standardization includes nomenclature for FEV1 or FVC measurements that do not fully meet the described criteria, but where quality limitations are considered to pose little risk of underestimation or overestimation of the measured value. For example, if the FVC from a single maneuver has a normal value but does not meet the indicators of adequate completion, it could be classified as usable. In this case, it can be assumed that it is unlikely that this measurement would be below the lower limit of normality if it had met the proper completion criteria. However, since the FVC could have a higher value if completion criteria were met, the FEV1/FVC ratio may be overestimated, so it will only be useful if it is decreased, as it could only be lower if the FVC reached a higher value by meeting completion criteria.
Quality grading of FEV1 and FVC.
| Quality grade | Number of measurements | Repeatability | Quality grade |
|---|---|---|---|
| A | ≥3 acceptable | Within 0.150L | Within 0.100La |
| B | 2 acceptable | Within 0.150L | Within 0.100La |
| C | ≥2 acceptable | Within 0.200L | Within 0.150La |
| D | ≥2 acceptable | Within 0.250L | Within 0.200La |
| E | ≥2 acceptable | >0.250L | >0.200La |
| U | 0 acceptable, 1 usable | N/A | N/A |
| F | 0 acceptable, 0 usable | N/A | N/A |
Quality grading is performed using the number of acceptable measurements and repeatability between the two best measurements of FEV1 and FVC, pre-bronchodilator and post-bronchodilator separately.
This classification system has proven useful in both epidemiological studies and clinical practice, having been implemented automatically in some spirometers and referenced in numerous publications [38–41]. However, it should be noted that in approximately 10–20% of cases, it is not possible to obtain maneuvers of good quality despite the technician's effort and the patient's good cooperation [40,41].
Reference valuesThe interpretation of spirometry is based on comparing the patient's measured values with the expected theoretical values for a healthy individual with similar demographic characteristics, allowing identification of deviations from normality that may indicate pulmonary pathology.
Therefore, it is essential to accurately record the following demographic data of the patient:
- •
Age: Should be calculated from the date of birth and the date of the test and expressed in years with one decimal.
- •
Height: Expressed in centimeters with one decimal. Measurement is performed without shoes, with feet together and back straight against a stadiometer. In cases of inability to stand or significant chest deformity, the patient's height can be measured by arm span (distance between the third finger of each hand with arms outstretched), or by measuring the heel-knee distance (applying a reference equation), recognizing differences by sex, age, and ethnicity in these estimations [42].
- •
Weight: Expressed in kilograms without decimals, rounded to the nearest 0.5kg. Measured with the patient without shoes. Using weight and height, the Body Mass Index (BMI) is calculated, expressed as kg/m2.
- •
Biological sex: Should be included in the spirometry request form, and if not, it should be requested from the patient at the time of the test. Biological sex is the determinant of predicted lung size, crucial for reference values.
- •
Race: The races currently used for GLI reference values are Caucasian (European ancestry), African American, Northeast Asian, Southeast Asian, and other/mixed.
Currently, it is recommended to use the GLI reference equations for all ages (3–95 years) [12]. These equations are based on a large international cohort (97,759 spirometries from 72 centers in 33 countries) and are recognized for minimizing age and ethnic biases. The adoption of GLI 2012, instead of the SEPAR reference equations proposed by García-Río [1], represents a significant improvement in standardization and international and multi-ethnic applicability. The use of previous SEPAR reference equations is only indicated if there is a very specific clinical or research justification for certain subpopulations, and must always be documented in the test report (Table 11).
Reference equations for spirometry values.
| García Rio 2013 [1] | GLI 2012 [44] | ||||||
|---|---|---|---|---|---|---|---|
| 6–20 years (Casan P) [45] | 20–65 years (Castellsaguer J) [46] | 65–85 years (García-Rio F) [47] | |||||
| Male | Female | Male | Female | Male | Female | ||
| FVC (L) | 0.02800T+0.03451P+0.05728E−3.21 | 0.03049T+0.02220P+0.03550E−3.04 | 0.0678T−0.0147E−6.0548 | 0.0454T−0.0211E−2.8253 | 0.0001572T2−0.00000268E3+0.223 | 0.0003171T2−0.0351E−6.368BSA+0.05925P+3.96 | log(Y)=a+b log(H)+c log(A)+age-spline+d group |
| FEV1 (L) | 0.02483T+0.02266P+0.07148E−2.91 | 0,02483T+0.02266P+0,07148E−2.9 | 0,0514T−0.0216E−3.9548 | 0,0326T−0.0253E−1.2864 | 0.0001107T2−0.0445E+2.886 | 0.0001726T2−0.0326E−2.303BSA+0.00012P2+3.398 | |
| %FEV1/FVC | 0.038T+0.140E−4.33 | 0.046T+0.051E−4.30 | −0.1902E+85.58 | −0.244E−0.1126P+94.88 | −0.0198E2+87.472 | −0.155T−0.184E+116.096 | |
GLI: global lung function initiative; FVC: forced vital capacity; FEV1: forced expiratory volume in the first second; %FEV1/FVC: ratio between FEV1 and FVC multiplied by 100; T: height (cm), P: weight (kg), E: age (years), BSA: body surface area (m2); Y: spirometric parameter, H: height (cm), A: age (years). ‘a’, ‘b’, ‘c’, ‘d’: coefficients that vary by ethnic group. ‘age-spline’: complex function requiring lookup tables or software implementation, especially for ages 3–25 years. ‘group’: Ethnicity (Caucasian, African-American, Northeast Asian, Southeast Asian, or Other).
For the interpretation of spirometry, it is necessary to include the patient's values, as well as the following data:
- •
Percentage of reference value (%ref.): The obtained values are commonly expressed as a percentage of the reference value, although using a fixed threshold as the lower limit of normality lacks statistical basis and may lead to classification errors.
- •
Lower limit of normality (LLN): A value is considered below the expected range if it is less than the 5th percentile of the reference population. The LLN should be calculated individually for each spirometry parameter according to the patient's sex, age, and height, providing a more accurate threshold.
- •
z-Scores: Expressing results as z-scores (z=(observed value−population mean)/standard deviation) is the most accurate and recommended method for interpreting functional abnormalities and their changes. The z-score allows standardized interpretation, where a z-score of −1.645 is equivalent to the 5th percentile, providing a value proportional to the standard deviation.
By adopting GLI and z-scores, misclassification of patients (for example, overdiagnosis of obstruction in the elderly or underdiagnosis in young adults, or incorrect interpretation of bronchodilator response) can be significantly minimized. This leads to more precise diagnoses, more appropriate initiation of therapies, and more reliable monitoring of disease progression over time and across different populations.
The current recommendation is to include all spirometry information, both values and graphs, in a single document. This document should also include the patient's demographic data, as well as the technician's observations and interpretation (Fig. 2) [43].
Recommended report of forced spirometry with bronchodilator test and slow spirometry. GLI: global lung function initiative. BMI: body mass index. FVC: forced vital capacity. FEV1: forced expiratory volume in 1second. %FEV1/FVC: ratio between FEV1 and FVC multiplied by 100. FET: forced expiration time. FIVC: forced inspiratory vital capacity. LLN: lower limit of normal. %ref.: % with respect to the mean of GLI reference values. SVC: slow vital capacity. IC: inspiratory capacity. ULN: upper limit of normal.
*Adapted from Culver BH et al. Am J Respir Crit Care Med 2017: 196, 1463–72. DOI: 10.1164/rccm.201710-1981ST [43].
The interpretation of spirometry should be clear, concise, and informative to help determine whether the observed result is within the normal range and, if not, what type of physiological abnormality is likely involved. Additionally, repeated spirometry assessment is important to detect deviations that may be clinically significant. Spirometry evaluates the flow and volume of expired air, with the most important indices being FEV1, FVC, and the FEV1/FVC ratio.
In spirometry, we must qualify and quantify the pattern:
- 1.
Airflow limitation or obstruction.
Obstructive ventilatory impairment is defined by an FEV1/FVC ratio below the lower limit of normality (LLN), which corresponds to the 5th percentile of a healthy population (Fig. 3) [6].
Ventilatory deficiency defined by spirometry: diagnostic algorithm. LLN (lower limit of normal) based on GLI 2012 reference equations [44].
The widely used cut-off points of 80% of the theoretical value for FEV1 (as it is arbitrary) and 0.70 for the FEV1/FVC ratio (as it does not account for age) are discouraged [44,48–50].
To evaluate the severity of pulmonary function impairment, cut-off points based on z-score values for FEV1 are recommended [51]. A z-score value between −1.645 and +1.645 is within the normal range, between −1.645 and −2.5 reflects mild obstructive impairment, between −2.5 and −4 moderate obstructive impairment, and z-score<−4 severe obstructive impairment [6].
Although the use of LLN and z-score is recommended for functional assessment, their application has so far been documented mainly in studies on obstructive pathology.
- 2.
Suspicion of restrictive pattern.
A restrictive disorder should be suspected when FVC is below the LLN, the FEV1/FVC ratio exceeds its LLN, and the flow-volume curve shows a convex morphology. However, this can only be confirmed if a reduction in total lung capacity (TLC) is objectively demonstrated, i.e., if TLC is below the LLN or the 5th percentile, by measuring static lung volumes [6,52,53].
- 3.
Unspecified ventilatory pattern and PRISm.
The presence of a pattern with reduced FVC and FEV1, with FEV1/FVC and TLC within the reference range, is known as an “unspecified” ventilatory impairment. Three-year follow-up of subjects with an “unspecified” pattern has shown that two-thirds maintain the alteration, while the remaining third develop either an overt obstructive or restrictive pattern [54]. When TLC is not available, the “unspecified” pattern is termed “preserved ratio impaired spirometry (PRISm)” [55].
- 4.
Suspicion of mixed ventilatory impairment.
A mixed ventilatory impairment is suspected when spirometry shows both FVC and FEV1/FVC below the LLN. To confirm the restrictive component, measurement of TLC, which must be below its LLN, is essential [6,56].
- 5.
Dysanapsis. Other measurements.
Dysanapsis is characterized by an FEV1/FVC ratio below the LLN, FEV1 within the reference range, and elevated FVC. This pattern may result from disproportionate growth between the pulmonary parenchyma and airways. Although it is believed to be a physiological variant of normality, new data suggest it may be associated with, or represent an early stage in the development of airway obstruction [57,58].
For the evaluation of peripheral airways, especially when FEV1 and FEV1/FVC are within the normal range, traditional measurements such as forced expiratory flow between 25% and 75% of FVC (FEF25–75) have been used; however, these measurements are highly variable, poorly reproducible, and lack specificity for small airway disease [59].
- 6.
Central airway obstruction.
Central and upper airways are located at the intrathoracic and extrathoracic levels. Obstruction of these airways in early stages may not cause a significant reduction in FEV1 or FVC, but often leads to a marked decrease in peak expiratory flow (PEF) relative to FEV1, as well as a reduction in forced inspiratory flow at 50% (FIF50%). The indices presented in Table 12 can help differentiate intrathoracic from extrathoracic obstruction [60,61], as well as the interpretation of flow-volume curves (Fig. 4).
Lung function indices useful for differentiating extrathoracic from intrathoracic obstructions.
| Extrathoracic obstruction | Intrathoracic obstruction | ||
|---|---|---|---|
| Fixed | Variable | ||
| PEF | Decreased | Normal or decreased | Decreased |
| FIF50% | Decreased | Decreased | Normal or decreased |
| FIF50%/FEF50% | ≈1 | <1 | >1 |
PEF: peak expiratory flow. FIF50%: forced inspiratory flow at 50% of FVC. FEF50%: forced expiratory flow at 50% of FVC. FVC: forced vital capacity.
It consists of measuring changes in pulmonary function after inhaling a short-acting bronchodilator. It helps diagnose and assess asthma, COPD, and other bronchial diseases, but the response to the bronchodilator test (BDT) is not sufficient to discriminate between asthma and COPD, nor to assess the definitive efficacy of a bronchodilator [6,62–67]. BDT should not be routine when baseline spirometry is normal or when the patient has previous BDT. It should be assessed based on available clinical information.
Indications:
- 1.
Baseline obstructive or mixed spirometry for the first time [6,49,68].
- 2.
Unspecified spirometric pattern in an individual with apparent good cooperation [6,49,68].
- 3.
Evaluation of bronchodilator response during an asthma attack, including bronchoconstriction induced in bronchial provocation tests [69].
- 4.
Disability assessment when FEV1 is below 65% of predicted [70,71].
- 5.
Preoperative evaluation when there is airflow limitation [72].
Contraindications:
- 1.
Known or probable adverse reactions to the bronchodilator to be used.
- 2.
Those of spirometry.
Bronchodilator drugs should be withdrawn beforehand (Table 5). If preparation has not been met, the test should be postponed. The relative merits of different protocols are unclear; the choice of bronchodilator, dose, and mode of administration is up to the laboratory [4,6,73]. Our recommendation is the use of salbutamol via MDI with a spacer. In children under 5 years, a mask is recommended. Unless contraindicated (known arrhythmias, tremor, or previous adverse reactions), high doses are recommended: 4 inhalations (400μg) separated by 30-s intervals, to minimize variability [4,6,74]. After 15min following salbutamol inhalation, the second spirometry should be performed 4.
Although the use of ipratropium bromide (IB) requires a longer waiting time (30min) and is a less effective drug for demonstrating reversibility in asthma, its use may be accepted in patients who may have intolerance or contraindications to beta-agonists.
Response variablesBoth FEV1 and FVC have been shown to be reproducible if the test is performed by well-trained personnel [6,12,49].
It is recommended to express the change in FEV1 or FVC as a percentage of predicted values, as this normalizes the result for sex, initial FEV1, and age, factors known to influence the response. A response is considered positive if the percentage change exceeds 10%. This criterion is supported by two major studies and is recommended by most respiratory societies [6,64,66,67,75–87].
% Change = (value after bronchodilators - value before bronchodilators) /predicted value × 100.
Regarding pediatric criteria, there are insufficient data to establish recommendations. The test could be considered positive if the percentage change is equal to or greater than 12% compared to baseline or 9–10% compared to predicted. To maintain adequate consistency, one criterion should be chosen and always applied [88–94]. Demonstration of reversibility by measuring PEF is useful for asthma diagnosis and outpatient management and for asthma attacks [95], but it is not recommended in pulmonary function labo
Artificial intelligence involvementNone of the material has been partially or totally produced with the help of any artificial intelligence software or tool.
Funding of the researchThis research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.
Conflicts of interestThe authors declare not to have any conflicts of interest that may be considered to influence directly or indirectly the content of the manuscript.






















